Skip to content
Radiation Lab RadiMeter
Test methods of IEC SC 45B standards

Test procedures

Test procedures by standard

A test procedure belongs to a standard: the clause numbers, the test points and the acceptance limits are all set by it. 8 procedures are written up so far, grouped below by the standard that specifies them. Each covers what the test is for, how to set it up, what the acceptance rule requires, and the mistakes that quietly invalidate a result — in our own words, with clause numbers cited for reference.

Scope of the procedures

The procedures are type tests: they establish how a model of instrument behaves across the conditions it will meet in use, rather than checking that one unit still reads correctly. Each write-up follows the same order — the purpose of the test, the set-up and test points, the acceptance rule, and the conditions that quietly invalidate a result — so they can be read side by side.

Procedure selection

Start with the three marked as core tests: statistical fluctuation, linearity and energy response. They describe how the instrument reads in the fields it was designed for, and the other results are hard to interpret without them. The cards marked optional cover angular response, dose rate dependence, overload, response time and alarm accuracy, which matter more or less depending on how the instrument is used, for example near pulsed X-ray equipment or as an area alarm.

Related standards

The written procedures follow IEC 60846-1:2009 for portable survey meters, dose rate meters and area monitors, the standard the evaluation tool implements. Reference fields are those of the ISO 4037 series. The other standards in the same series are listed at the end of this page; each has its own page setting out its scope and the characteristics it requires.

IEC 60846-1:2009

Tool available 8 procedures

Portable survey meters, dose rate meters and area monitors

Standards without written procedures

These are described on the site, but their test procedures are not written up yet. The standard page for each covers its scope and the characteristics it requires.

Version 1.1.0 · 29 October 2026 · What changed