Skip to content
Radiation Lab RadiMeter
Test methods of IEC SC 45B standards

About

RadiMeter documents the IEC standards for radiation measuring instruments and provides free browser tools that perform their characteristic evaluation — the type test that checks whether an instrument meets the performance requirements of a standard. It describes 8 standards across the instrument classes of IEC subcommittee SC 45B, and adds evaluation tools standard by standard: the standard's own page first, then its test procedures, then the tool. The tools page lists which evaluations run today and what the other standards need first.

Intended users

People who put instruments in front of a source: instrument manufacturers running acceptance checks, radiation safety officers verifying that a meter still behaves as specified, and radiation workers who need a defensible record of what an instrument did. The calculations are the ones such teams usually keep in a spreadsheet, with the arithmetic fixed and checked.

Origin

The calculation engine was written to replace a working spreadsheet that a testing team used for linearity, statistical fluctuation and energy response checks. The starting requirement was blunt: it had to reproduce that spreadsheet's results to the last decimal place, or nobody would trust it.

Three things were later changed on purpose, so the engine no longer matches the spreadsheet everywhere. Statistical fluctuation is judged by the method of clause 8.7.4, at every test point against a limit that steps with dose rate, rather than from one coefficient of variation at a single point against a fixed 5%. Background is subtracted from the readings automatically, where the spreadsheet expected readings with the background already removed. And means keep one significant figure more than the readings, where the spreadsheet rounded them to a fixed four. Defects found in a hidden, unused part of the workbook were never carried over. The changes page records these changes, made while the engine was being written, and every change since.

The requirement to reproduce the spreadsheet shaped the design. The engine is a set of pure functions with no dependence on the interface around them, and it is verified against a fixed set of reference vectors on every change, so results cannot drift quietly between versions. The same engine runs behind every evaluation on this site.

Coverage

The written material covers each standard, the instrument classes it applies to, the test procedures it specifies, and the reference topics that apply whichever standard you work to — operational quantities, reference radiation qualities, measurement uncertainty. Scope, acceptance criteria and clause numbers belong to each standard, so they are stated on that standard's own page rather than here.

Where an evaluation tool exists, it computes each test point, applies the acceptance criteria of the standard and reports a verdict per test and overall. It also computes the combined standard uncertainty of each result and carries it through to the relative response. That uncertainty is reported, not applied to the decision rule — see the Disclaimer for why the distinction matters.

Data handling

The calculations run in the browser, and the Excel export is generated locally. Using the tool needs a Google sign-in, and the sessions, presets and acceptance criteria you save, instrument identifiers and readings included, are stored in your Radiation Lab account so that they follow you between devices. The standards, instrument and guide pages need no account. The full detail is on the Privacy page.

Maintainer

RadiMeter is built and maintained by one person, who works in the radiation field at a non-profit organisation. It is a personal project, built outside that work: it is not published by that organisation, carries no institutional endorsement, and the organisation is not responsible for anything here. Questions, corrections and reports of a wrong result reach the same person directly — contact. A correction that changes a result or a verdict is released as a new version. Every page shows the current version above its footer, and Changes between versions lists what each version changed, computed values and verdicts first.

Independence

RadiMeter is an independent project. It is not affiliated with, endorsed by or connected to the IEC, the ISO, or any standards or accreditation body. It is not a testing laboratory — nothing is tested here on your behalf and no report is issued. It cites clause numbers and describes methods in its own words; it does not reproduce the text or tables of any standard, and it is not a substitute for obtaining a standard from its publisher.

Roadmap

The engine is verified against reference vectors on every change, and the written material covers the standards, their test procedures and the reference topics. Work continues on breadth: each further instrument class needs its own acceptance criteria and, in several cases, a measurement condition the engine does not model yet — irradiation on a phantom for active personal dosemeters, elapsed time to readout for passive systems, spectrum uncertainty for neutron instruments.

If a result looks wrong, or the instrument you need to test is not covered yet, that is exactly what this phase is for. Tell us.

Funding

It is free and there is no paid tier. There is no advertising on the site today; the intention is to fund it with contextual advertising later, which is why the Privacy page already says what will change when that happens.

Version 1.1.0 · 29 October 2026 · What changed